07c9332fdf224cc8c847f6d6963a3e4cadec86cc
[visomat-utils.git] / README
1 Utility software to access the Visomat Double Comfort blood pressure meter:
2 http://www.visomat.de/Oberarm-Blutdruckmes.372.0.html
3
4 The device can store measurements for two distinct users and has a USB port
5 that can be used to download the history of measurements or trigger a new
6 measurement (apparently only for user 1).
7
8 The device is seen by linux as a non-standard USB CDC device:
9
10   [10217.072073] usb 2-5: new full-speed USB device number 5 using ohci_hcd
11   [10217.292067] usb 2-5: New USB device found, idVendor=1247, idProduct=00f8
12   [10217.292079] usb 2-5: New USB device strings: Mfr=1, Product=2, SerialNumber=3
13   [10217.292087] usb 2-5: Product: USB COMMUNICATION BRIDGE
14   [10217.292093] usb 2-5: Manufacturer: JAPAN PRECISION INSTRUMENTS
15   [10217.301191] cdc_acm 2-5:1.0: Zero length descriptor references
16   [10217.301212] cdc_acm: probe of 2-5:1.0 failed with error -22
17
18 This could be fixed adding a quirk to the cdc_acm driver in linux, but there
19 is not much of a point for this as the data is downloaded using USB bulk
20 communication anyways.
21
22 Each data packet starts with the STX control character, and ends with ETX.
23
24 There are at least two kind of packets: requests and responses.
25
26 A request can be used to send command codes to the device.
27
28 EEPROM data request:
29   
30   R00
31
32 Date and time request:
33
34   R02
35
36 A response can be read after a request has been sent.
37
38 The payload of a response to a "R00" request is something like this:
39
40   M106100101120211511021201001011208115610711610010112030103065073440601132001130680661001011311012708406610010112030106068066M20710010112031155111116100101120201170700764406011312012807107244060113170126050065440601131901140630661001011400011407007113031211450121080072
41
42 The structure can be better visualized when split in fields:
43
44   M  # start of a Memory block
45   1  # memory block index, corresponding to a "User ID"
46   06 # number of records in this memory block
47
48   # Each records stores date, time, systolic and diastolic pressure, pulses
49
50   # yymmdd HHMM systolic diastolic pulses
51     100101 1202     1151       102    120
52     100101 1208     1156       107    116
53     100101 1203     0103       065    073
54     440601 1320     0113       068    066
55     100101 1311     0127       084    066
56     100101 1203     0106       068    066
57
58   # Same structure for the other memory block
59   M
60   2
61   07
62   100101 1203 1155 111 116
63   100101 1202 0117 070 076
64   440601 1312 0128 071 072
65   440601 1317 0126 050 065
66   440601 1319 0114 063 066
67   100101 1400 0114 070 071
68   130312 1145 0121 080 072