README: mention that the Prevention DS-2200PV uses the same protocol
[visomat-utils.git] / README
diff --git a/README b/README
index 10b889e..7021cd1 100644 (file)
--- a/README
+++ b/README
@@ -1,6 +1,8 @@
 Utility software to access the Visomat Double Comfort blood pressure meter:
 http://www.visomat.de/Oberarm-Blutdruckmes.372.0.html
 
 Utility software to access the Visomat Double Comfort blood pressure meter:
 http://www.visomat.de/Oberarm-Blutdruckmes.372.0.html
 
+The same protocol seems to be used also by the Prevention DS-2200PV.
+
 The device can store measurements for two distinct users and has a USB port
 that can be used to download the history of measurements or trigger a new
 measurement (apparently only for user 1).
 The device can store measurements for two distinct users and has a USB port
 that can be used to download the history of measurements or trigger a new
 measurement (apparently only for user 1).
@@ -68,3 +70,19 @@ The structure can be better visualized when split in fields:
   440601 1319 0 114 063 066
   100101 1400 0 114 070 071
   130312 1145 0 121 080 072
   440601 1319 0 114 063 066
   100101 1400 0 114 070 071
   130312 1145 0 121 080 072
+
+When the number of records is 0, there is a dummy byte at the end of the user
+memory block, e.g.:
+
+  M1001M20115110321300134097083
+
+The structure is:
+
+  M
+  1
+  00
+  1   # dummy byte, present only when the previous field is 00
+  M
+  2
+  01
+  151103 2130 0 134 097 083